A Low Background Gamma Ray Spectrometer with Anticosmic Shielding

Nguyen Quoc Hung, Vo Hong Hai, Tran Kim Tuyet, Ho Lai Tuan
Author affiliations

Authors

  • Nguyen Quoc Hung University of Science, HCMC-Vietnam National University
  • Vo Hong Hai University of Science, HCMC-Vietnam National University, 227 Nguyen Van Cu, Ward 4, District 5, Ho Chi Minh City, Vietnam
  • Tran Kim Tuyet University of Science, HCMC-Vietnam National University, 227 Nguyen Van Cu, Ward 4, District 5, Ho Chi Minh City, Vietnam
  • Ho Lai Tuan University of Science, HCMC-Vietnam National University, 227 Nguyen Van Cu, Ward 4, District 5, Ho Chi Minh City, Vietnam

DOI:

https://doi.org/10.15625/0868-3166/26/1/7474

Keywords:

Gamma ray spectrometer, cosmic rays, plastic scintillator, high-purity Germanium detector

Abstract

The article describes a gamma ray spectrometer protected by a lead shield (Model 747E Canberra lead shield) and an active shield made of an 80~cm \(\times\) 80~cm \(\times\) 3~cm plastic scintillator plate in anticoincidence on top of the lead shield. The detector used as low background gamma-spectrometer is a high purity Germanium crystal of model GC2018 Canberra. The background count rate currently achieved (30-2400 keV) is 1.27 cps without anticoincidence. The level of background suppression obtained from the active protection is 0.80 overall and about 0.43 for the 511 keV gamma line. The gamma ray spectrometer is installed and operated in the Nuclear Laboratory, Department of Nuclear Physics, University of Science, HCMC-Vietnam National University.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

References

G. Heusser, The background components of germanium low-level spectrometers, Nuclear Instruments and Methods in Physics Research B: Vol.17,1986, pp.418-422 DOI: https://doi.org/10.1016/0168-583X(86)90175-8

H.L. Malm and M.M. Watf, Background reduction in germanium spectrometers: Material selection geometry and shielding, Nuclear Instruments and Methods in Physics Research: Vol.223, 1984, pp.420-425. DOI: https://doi.org/10.1016/0167-5087(84)90686-0

Richard M. Lindstrom, David J. Lindstrom, Lester A. Slaback and John K. Langland, A low-background gamma-ray assay laboratory for activation analysis, Nuclear Instruments and Methods in Physics Research A: Vol.299, 1990, pp.425-429. DOI: https://doi.org/10.1016/0168-9002(90)90818-Q

Downloads

Published

27-07-2016

How to Cite

[1]
N. Q. Hung, V. H. Hai, T. K. Tuyet and H. L. Tuan, A Low Background Gamma Ray Spectrometer with Anticosmic Shielding, Comm. Phys. 26 (2016) 93. DOI: https://doi.org/10.15625/0868-3166/26/1/7474.

Issue

Section

Papers
Received 27-11-2015
Accepted 14-04-2016
Published 27-07-2016