Forthcoming

Determination of linear attenuation coefficients of Al and Pb for collimated 662 keV gamma rays using TCS-172 gamma scintillation survey meter

Linear attenuation coefficients of Al and Pb for collimated 662 keV gamma rays

Quyet Pham Dang, Trinh Thi Tu Anh
Author affiliations

Authors

  • Quyet Pham Dang Dalat https://orcid.org/0000-0002-8984-7747
  • Trinh Thi Tu Anh Office of National Assembly Delegations and People Councils, 02 Tran Hung Dao, Dalat, Vietnam

DOI:

https://doi.org/10.15625/0868-3166/22766

Keywords:

NaI(Tl), TCS-172, linear attenuation, 662 keV, gamma survey meter

Abstract

Linear attenuation coefficients of Al and Pb for the 662 keV gamma rays have been carried out at the Training Center of Dalat Nuclear Research Institute using the Cs-137 source and the handheld equipment TCS-172 gamma scintillation survey meter. The gamma-ray beam emitted from the Cs-137 source was collimated with a narrow beam geometry. The results of our investigation were compared and found to be in good agreement with previously published data.

Downloads

Download data is not yet available.

References

[1] J. E. Martin, Physics for Radiation Protection. Wiley-VCH, 2006, 10.1002/9783527618798.

[2] A. L. Conner, H. F. Atwater and E. H. Plassmann, Gamma-ray attenuation-coefficient measurements, Phys. Rev. A 1 (1970) 539.

[3] B. Buyuk and A. B. Tugrul, Comparison of lead and wc-co materials against gamma irradiation, Acta Phys. Pol. A 125 (2014) 423.

[4] S. Gjorgieva and L. Barandovski, Measurement of the mass attenuation coefficient from 81 kev to 1333 kev for elemental materials al, cu and pb, AIP Conf. Proc. 1772 (2016) 180003.

[5] Spectrum Techniques, LLC, Radiation absorber set - 20 piece, n.d.

[6] Board of Radiation and Isotope Technology, Government of India Department of Atomic Energy, Certificate of calibration, no. tt & cd/pkp/2000/1694, 2000.

[7] Aloka Co., Ltd., Gamma scintillation survey meter, Model TCS-172 instruction manual, n.d.

[8] T. E. Johnson, Introduction to Health Physics. McGraw-Hill Education, 5th ed., 2017.

[9] J. R. Taylor, An Introduction to Error Analysis: The Study of Uncertainties in Physical Measurements. University Science Books, 1997.

Published

26-11-2025

How to Cite

[1]Q. Pham Dang and T. T. A. Trinh, “Determination of linear attenuation coefficients of Al and Pb for collimated 662 keV gamma rays using TCS-172 gamma scintillation survey meter: Linear attenuation coefficients of Al and Pb for collimated 662 keV gamma rays”, Comm. Phys., vol. 35, no. 4, Nov. 2025.